User Manual

1. Adjusting the Sample Probe Height
2. Revive After Storage Routine (Luminex)
3. System Initialization
All system startup routines are run from the MAGPIX Off-Plate Reagent Block. For
detailed instructions on how to perform these operations, refer to the Starting the
MAGPIX System for the First Time section of this manual.
FIGURE 6.
MAGPIX Off-Plate Reagent Block
1 Reservoirs
2 Strip Wells
MAGPIX New System Startup Errors
Problem Summary Resolution Notes
Calibration Failure
(2x) Clean with 70% isopropanol
or ethanol alcohol (reservoir B)
(2x) Alcohol Flush with 70%
isopropanol or ethanol alcohol
(reservoir B)
(3x) Rinse (reservoir D)
(3x) Prime
Run Calibration Routine Verify that reagent is freshly
vortexed. Use 6 drops of
reagent.
MAGPIX
®
User Quick Guide
18
For In Vitro Diagnostic Use