Hardware User (IVD) Instruction Manual

11
CHAPTER 3 Installation Procedure
Before handling or unpacking MAGPIX, make certain that the selected site is appropriate.
See “Installation Diagrams” on page 28 for handling and site installation requirements and
detailed dimensions of MAGPIX.
Check for these requirements:
Indoors
Operating temperature of 15-35°C (59-95°F)
Operating relative humidity of 20-80%, noncondensing
Operating altitude up to 2400m (7874 ft) above mean sea level
Available electrical power outlet with protective earthing and easy accessibility
Available area of approximately 3‘ X 3’ (91.44cm), including a 2” (5.08cm) clearance
between the back of MAGPIX and any wall or vertical surface.
Stable, level surface
MAGPIX arrives in a large, corrug
ated cardboard overpack on skids.
FIGURE 9. The MAGPIX Overpack
CAUTION: This overpack is too heavy to be lifted by one person
(approximately 119lbs (53.97kg), a three-person lift) and should
be moved mechanically. Be careful the overpack is not punctured
during any necessary moving.