Luminex® LDS 1.7 SP2 Upgrade Instructions to xPONENT® 3.
© LUMINEX CORPORATION, 2001-2009. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into any language or computer language, in any form or by any means without prior express, written consent of: LUMINEX CORPORATION 12212 Technology Boulevard Austin, Texas 78727-6115 U.S.A. Voice: (512) 219-8020 Fax: (512) 219-5195 Luminex® LDS 1.7 SP2 Upgrade Instructions to xPONENT® 3.1 SP1 PN 89-00002-00-212 Rev.
Contents Upgrade Procedure .............................................................................1 Important Upgrade Information..........................................................2 Recommendations...............................................................................2 Training...............................................................................................3 System Requirements .........................................................................3 Kit Components ......
ii
Upgrade Procedure 1
Luminex LDS 1.7 SP2 Upgrade Instructions to xPONENT 3.1 SP1 - For In Vitro Diagnostic Use Important Upgrade Information Before performing the Luminex xPONENT 3.1 SP1 upgrade, verify that the following cables are in place on your system: • • • USB cable from P1 on the Luminex 100 or Luminex 200™ analyzer to a USB port on the PC. CANBUS cable from P1 on the Luminex SD to P2 on the analyzer. Serial cable from the Luminex XY Platform™ to the PC.
Luminex LDS 1.7 SP2 Upgrade Instructions to xPONENT 3.1 SP1 - For In Vitro Diagnostic Use Training System Requirements To register or find out more about Luminex xPONENT 3.1 SP1 training, contact Luminex’s Services Sales Team at serviceandtraining@luminexcorp.com or 512-219-8020, ext. 4386. Visit www.xmapuniversity.com for more details.
Luminex LDS 1.7 SP2 Upgrade Instructions to xPONENT 3.
Luminex LDS 1.7 SP2 Upgrade Instructions to xPONENT 3.1 SP1 - For In Vitro Diagnostic Use 7. Connect power cables for the monitor, computer, Luminex analyzer, and Luminex XYP. 8. If you are installing a touchscreen monitor, you will need to install the touchscreen drivers and calibrate the touchscreen according to manufacturer’s instructions, or contact Luminex Technical Support for assistance.
Luminex LDS 1.7 SP2 Upgrade Instructions to xPONENT 3.1 SP1 - For In Vitro Diagnostic Use Uninstall Luminex Software Note: If you are installing a new PC, you do not need to uninstall Luminex software. During the Luminex software removal, error dialog boxes may open. They do not affect the uninstall. Click OK or Yes to continue the uninstall process. If you are continuing from the Remove Luminex LMAT Software procedure, skip to step 4. To remove the software: 1. Close any open applications. 2.
Luminex LDS 1.7 SP2 Upgrade Instructions to xPONENT 3.1 SP1 - For In Vitro Diagnostic Use Note: At this point in the installation (step 5), the PC may reboot. If the InstallShield wizard fails to appear after rebooting, ensure the user is logged in as admin or as a user with installation privileges, and manually run (setup.html) again from the DVD. To run the setup file manually, open the PC directory structure and located the DVD drive. Open the directory on that drive. Locate and click set.up.html.
Luminex LDS 1.7 SP2 Upgrade Instructions to xPONENT 3.1 SP1 - For In Vitro Diagnostic Use Verify Firmware Version Verify that the correct firmware version for the Luminex analyzer and the Luminex XYP display in the xPONENT software. Click Maintenance>System Info. The Detector Firmware Version (Luminex 100 or Luminex 200) and the XY Firmware Version fields should display the latest firmware version numbers. • If you have a Luminex 100, the firmware version should be 2.3.9.