Print Batch Worklist Prints the status of each of the commands in the command list, as
well as the microtiter plate graphic.
Print Report Prints the data interpretation report including the standards, controls,
graph of standards, and Samples data.
Replay Batch Replays batch sample acquisition. All batch commands are replayed
allowing for batch acquisition viewing exactly as in real-time. You
cannot replay New Advanced Batches. When running a Replay
Batch command, the original acquisition data is not altered. A new
file is created. The most common use is for system demonstration
and training.
Report Raw Fluorescence Enables the median fluorescence intensity (MFI) to appear on the
Analyte Report.
Resume Resumes the process that was paused.
Sample Probe Down Raises and lowers the sample probe to adjust the probe height. This
adjustment is recommended when using different microtiter plates or
when changing sample probes. Refer to the Luminex 200 System
Manual or the Luminex 100 IS User Manual for Version 2.3 for the
adjusting procedure.
Sanitize Uses the Luminex XYP reservoir due to volume requirement. The
sanitize command performs a similar function as the alcohol flush
command. However, the sanitize command uses 10% to 20%
household bleach to decontaminate sample lines and the cuvette after
biohazard contact.
Save and Load If selected, displays the run batch tab immediately after you set up
batch information and click Finish.
Save Only Allows you to save the batch information. User can open the batch
later to acquire data.
Self Diagnostics Performs a self-diagnostics to see if the Luminex 200 analyzer and
all system operations are functioning correctly. Self-Diagnostics tests
these functions:
Flash memory test
Microcontroller RAM test
Nonvolatile memory test
DSP program CRC test