Software

xMAP Technology System Overview
PN 89-00002-00-150 Rev. A 3 - 3
Sensitivity Detect 1000 fluorochromes phycoerythrin (PE) per xMAP
microsphere
Reporter channel dynamic range: 3.5 decades of detection
Capacity The specifications below reflect minimum capacity values:
Distinguish a minimum of 1 to a maximum of 100 unique xMAP
microsphere sets in a single sample
Detect and distinguish surface reporter fluorescence emissions at
575 nm on the surface of 1-100 unique xMAP microspheres sets
in a single sample
Sample core: 15-20 µm core at 1 µL/sec. sample inject rate
Maintain samples at a constant temperature from 35°C to 55°C
(95°F to 131°F)
Automatic sampling from a 96-well plate
Start sampling from any well position
Sheath container and waste container hold enough volume to run
up to two 96-well plates between refills
Microtiter plates with 96 wells must be compatible with the
Luminex XYP instrument plate holder. The following microtiter
plate types are compatible with the Luminex XYP instrument
plate holder: any color of flatbottom, conical, round, filter
bottom, or half-well plates. The overall plate must not have an
overall height of more than 0.75” (19 mm).
Microtiter plates with 96 wells must be compatible with
Luminex XYP instrument heater block temperature from 35°C to
55°C (95°F to 131°F) when performing heated assays and using
the heater block.
Luminex 200
Analyzer
Specifications
Indoor use only
Operating temperature: 15°C to 30°C (59°F to 86°F)
Humidity: 20% to 80%, noncondensing
Altitude: Operation up to 2400 m (7874 ft.) above mean sea level
Physical dimensions: 43 cm (17 inches) W x 50.5 cm (20 inches)
D x 24.5 cm (9.5 inches) H
Weight: maximum of 25 kg (60 lbs.)
UL installation category: UL Installation Category II, as defined
in Annex J of UL 61010A-1
Pollution degree: UL Pollution Degree 2, as defined in Section
3.7.3.2 of UL 61010A-1
Shipping and storage: The allowable shipping and storage
temperature and humidity ranges are 0°C to + 50°C and 20-80%
noncondensing, respectively
Input voltage range: 100 - 120 V~ ± 10%, 1.4 Amp, and 200-240
V~ ± 10%, 0.8 Amp, 47-63 Hz.