Software Owner manual

xMAP Technology The System
PN 89-00002-00-144 Rev. B 2 - 3
Required
Laboratory
Reagents
Household bleach
70% isopropanol or 70% ethanol
Mild detergent
Distilled water
Luminex IS 2.3
Software
Luminex IS 2.3 software provides complete control of the system
and performs data analysis. Your Luminex 200 system is preloaded
with the Luminex software. However, we supply a software CD
should you need to reinstall the software.
This software requires a dedicated system. Unauthorized additional
software is prohibited and may result in improper operation of the
system.
Luminex 200
Performance
Specifications
Speed Speed: greater than or equal to1.7 GHz Intel® Pentium® IV
processor with 256 MB RAM
USB communications link for fast data transfer
Automatic transfer of assay templates and new reagent
information into the system via a large capacity read/write CD
Installation: < 4 hours
System calibration: < 10 minutes
System controls: < 10 minutes
Barcode reader entry of sample IDs
Automatic post-analysis
Analyze one 96-well plate/hour depending on manufacturer’s kit
Up to 100 xMAP microsphere sets per sample
System warmup: 30 minutes. Systems that remain inactive for at
least four hours will require a warm-up to restart the lasers. After
acquiring sample, running system calibrators, running system
controls, and warming up the instrument, the system resets the
four-hour internal clock.
Accuracy and Precision Sample uptake volume: ± 5%
Classification of xMAP microspheres: > 80%
Misclassification of xMAP microspheres: 2% - may vary by
xMAP microsphere product lines. Refer to the specific product
information sheet for further details.
Temperature control: 0°C to + 2°C of target