User Manual User Manual

Shipping and storage: The allowable shipping and storage temperature and humidity
ranges are 0°C to + 50°C and 20-80% noncondensing, respectively
Input voltage range: 100 - 120 V~ ± 10%, 1.4 Amp, and 200-240 V~ ± 10%, 0.8 Amp,
47-63 Hz.
AC inlet fuse: 3 Amp, 250 V~, fast acting
Optics
Reporter laser: 532 nm, nominal output 10-15 mW, maximum 500 mW, frequency-doubled
diode; mode of operation, continuous wave (CW)
Classification laser: 635 nm, 9.1 mW ± 6%, maximum output 25 mW, diode; mode of
operation, continuous wave (CW)
Reporter detector: Photomultiplier tube, detection bandwidth of 565-585 nm
Classification detector: Avalanche photo diodes with temperature compensation
Doublet discrimination detector: Avalanche photo diodes with temperature compensation
Fluidics
Sheath flow rate 90 µL ± 5 µL/second
Cuvette: 200 micron square flow channel
Sample injection rate: 1 µL/second ± 0.05 µL
Sample uptake volume: 20-200 µL
Electronics
Reporter channel detection: A/D resolution 14 bits
Communications interface: USB
Luminex XYP instrument, communications interface: RS 232
Luminex XYP Instrument General
Ambient temperature: 15°C to 30°C (59°F to 86°F)
Humidity: 20% to 80%, noncondensing
Altitude: operation up to 2400 m (7874 ft) above mean sea level
Physical dimensions: 44 cm (17.25 inches) W x 60 cm (23.5 inches) D x 8 cm (3 inches) H
Weight: 15 kg (33 lbs.)
UL installation category: UL Installation Category II, as defined in Annex J of UL 61010A-1
Pollution degree: UL Pollution Degree 2, as defined in Section 3.7.3.2 of UL 61010A-1
Heater operating range: 35°C to 55°C (95°F to 131°F) with tolerance 0°C to +2°C
Input voltage range: 100-240 V~ ± 10%, 1.8 Amps, 47-63 Hz
AC inlet fuse: 3 A, 250 V~, fast acting
For In Vitro Diagnostic Use
The System
13