User Manual User Manual

Required Laboratory Reagents
Household bleach
70% isopropanol or 70% ethanol
Mild detergent
Distilled water
Luminex Software
Luminex software provides complete control of the system and performs data analysis. Your
Luminex 200 system is preloaded with the Luminex software. However, we supply a software
DVD should you need to reinstall the software.
This software requires a dedicated system. Unauthorized additional software is prohibited
and may result in improper operation of the system.
Luminex 200 Performance Specifications
Speed
USB communications link for fast data transfer
Automatic transfer of assay templates and new reagent information into the system via a
lare capacity read/write DVD
Installation: < 4 hours
System calibration: < 10 minutes
System controls: < 10 minutes
Barcode reader entry of sample IDs
Automatic post-analysis
Analyze one 96-well plate/hour depending on manufacturer’s kit
Up to 100 xMAP microsphere sets per sample
System warmup: 30 minutes. Systems that remain inactive for at least four hours will
require a warm-up to restart the lasers. After acquiring sample, running system calibrators,
running system controls, and warming up the instrument, the system resets the four-hour
internal clock.
Accuracy and Precision
Sample uptake volume: ± 5%
Classification of xMAP microspheres: > 80%
Misclassification of xMAP microspheres: ≤ 2 % - may vary by xMAP microsphere product
lines. Refer to the specific product information sheet for further details.
Temperature control: 0°C to + 2°C of target
For In Vitro Diagnostic Use
The System
11