User Manual Version 1.7 User guide

xMAP Technology System Overview
PN 89-00002-00-063 Rev. A 3 - 3
Laser Reporter laser: 532 nm, nominal output 10-16.5 mW, maximum
500 mW, frequency-doubled diode; mode of operation mW,
continuous wave (CW)
Classification laser: 635 nm, nominal output 9.1 mW ± 6%,
maximum output 25 mW, diode; mode of operation, continuous
wave (CW)
Fluidics Sheath flow rate: 90 µL/second ± 5 µL/second
Cuvette: 200 micron square flow channel
Sample injection rates: 60 µL/minute, 45 µL/minute, or 30 µL/
minute
Sample uptake volume: 10-200 µL
Electronics Reporter channel detection: photomultiplier tube,
A/D resolution, 14 bits
Classification and doublet discriminator channel detection:
avalanche photo diodes (APD) with temperature
compensation, A/D resolution 12 bits
Communications interface: RS 232
Signal Processing Signal modes: linear, with logarithmic or linear display option
PC Computer: a Pentium III microprocessor or equivalent class PC,
minimum speed 933 MHz
Main memory: minimum of 128 MB
Hard disk drive: 10 GB minimum storage capacity
Communications, serial ports: minimum of 2 RS-232 compatible
serial ports
Operating system support: MS
Windows
®
CE marked and UL listed
Two-button mouse or equivalent
Monitor: minimum of 17-inch diagonal monitor
Keyboard: a 104-key keyboard or equivalent
Power cords for the US and Canada are included in the system.
Power cords for other countries must be ordered separately.
Power: 115-230 V~, Maximum 6.0 Amps, 50-60 Hz