User Manual Version 1.7 User guide

Luminex 100 User Manual Version 1.7 xMAP Technology
3 - 2 PN 89-00002-00-063 Rev. A
In addition, you can purchase the optional Luminex Sheath Delivery
system (Luminex SD).
xMAP Reagents • xMAP classification calibration microspheres (CAL1)
• xMAP reporter calibration microspheres (CAL2)
• Sheath fluid
Software Luminex Data Collector software provides complete control of the
Luminex 100 analyzer and performs real-time digital analysis of
assays. The Luminex 100 system is preloaded with the Luminex
software. However, we supply a software CD in case you need to
reinstall the software.
Luminex 100
Analyzer
Specifications
General • Operating temperature: 15°C to 30°C (59°F to 86°F)
• Storage temperature: 0°C to 50°C (32°F to 122°F)
• Humidity: 20% to 80%, noncondensing
• Altitude: designed to operate up to 2400 m (7874 ft.) above
mean sea level
• Physical dimensions: 43 cm (17 in) W x 51 cm (20 in) D x 23 cm
(9 in) H
• Weight: 23 kg (60 lbs)
• UL installation category: UL Installation Category II, as defined
in Annex J of UL 61010A-1
• Pollution degree: UL Pollution Degree 2, as defined in section
3.7.3.2 of UL 61010A-1
Power • Input voltage range:
• 100-240 V~, 1.5 Amps, 47-63 Hz
OR
• 100-120 V~ and 200-240 V~, 1.4 Amps, 47-63 Hz. See label
on rear of instrument.
• AC inlet fuse: 3A, 250V~, Fast Acting
OR
2A, 250 V~, Fast Acting. See label on rear of instrument.
• Shipping and storage: allowable shipping and storage
temperature and humidity ranges, 0°C to +50°C and 20%-80%
noncondensing, respectively.