User Manual Version 1.7 User guide

Luminex 100 User Manual Version 1.7 xMAP Technology
3 - 2 PN 89-00002-00-063 Rev. A
In addition, you can purchase the optional Luminex Sheath Delivery
system (Luminex SD).
xMAP Reagents xMAP classification calibration microspheres (CAL1)
xMAP reporter calibration microspheres (CAL2)
Sheath fluid
Software Luminex Data Collector software provides complete control of the
Luminex 100 analyzer and performs real-time digital analysis of
assays. The Luminex 100 system is preloaded with the Luminex
software. However, we supply a software CD in case you need to
reinstall the software.
Luminex 100
Analyzer
Specifications
General Operating temperature: 15°C to 30°C (59°F to 86°F)
Storage temperature: 0°C to 50°C (32°F to 122°F)
Humidity: 20% to 80%, noncondensing
Altitude: designed to operate up to 2400 m (7874 ft.) above
mean sea level
Physical dimensions: 43 cm (17 in) W x 51 cm (20 in) D x 23 cm
(9 in) H
Weight: 23 kg (60 lbs)
UL installation category: UL Installation Category II, as defined
in Annex J of UL 61010A-1
Pollution degree: UL Pollution Degree 2, as defined in section
3.7.3.2 of UL 61010A-1
Power Input voltage range:
100-240 V~, 1.5 Amps, 47-63 Hz
OR
100-120 V~ and 200-240 V~, 1.4 Amps, 47-63 Hz. See label
on rear of instrument.
AC inlet fuse: 3A, 250V~, Fast Acting
OR
2A, 250 V~, Fast Acting. See label on rear of instrument.
Shipping and storage: allowable shipping and storage
temperature and humidity ranges, 0°C to +50°C and 20%-80%
noncondensing, respectively.