User Manual, Version 2.3

xMAP Technology The System
PN 89-00002-00-071 Rev. C 3 - 5
55°C (95°F to 131°F) when performing heated assays and using
the heater block
Luminex 100
Analyzer General
Indoor use only
Operating temperature: 15°C to 30°C (59°F to 86°F)
Humidity: 20% to 80%, noncondensing
Altitude: Operation up to 2400 m (7874 ft.) above mean sea level
Physical dimensions: 43 cm (17 inches) W x 50.5 cm (20 inches)
D x 24.5 cm (9.5 inches) H
Weight: maximum of 25 kg (60 lbs.)
UL installation category: UL Installation Category II, as defined
in Annex J of UL 61010A-1
Pollution degree: UL Pollution Degree 2, as defined in Section
3.7.3.2 of UL 61010A-1
Shipping and storage: The allowable shipping and storage
temperature and humidity ranges are 0°C to + 50°C and 20-80%
noncondensing, respectively
Input voltage range: 100 - 120 V~ ± 10%, 1.4 Amp, and 200-240
V~ ± 10%, 0.8 Amp, 47-63 Hz.
AC inlet fuse: 3 Amp, 250 V~, fast acting
Optics Reporter laser: 532 nm, nominal output 10-15 mW, maximum
500 mW, frequency-doubled diode; mode of operation,
continuous wave (CW)
Classification laser: 635 nm, 9.1 mW ± 6%, maximum output 25
mW, diode; mode of operation, continuous wave (CW)
Reporter detector: Photomultiplier tube, detection bandwidth of
565-585 nm
Classification detector: Avalanche photo diodes with temperature
compensation
Doublet discrimination detector: Avalanche photo diodes with
temperature compensation
Fluidics Sheath flow rate 90 µL ± 5 µL/second
Cuvette: 200 micron square flow channel
Sample injection rate: 1 µL/second
Sample uptake volume: 20-200 µL
Electronics Reporter channel detection: A/D resolution 14 bits
Communications interface: USB
Luminex XYP instrument, communications interface: RS 232