User Manual, Version 2.1 User guide
xMAP Technology The System
PN 89-00002-00-070 Rev. A 5 - 5
Luminex 100
Analyzer
General • Operating temperature: 15°C to 30°C (59°F to 86°F)
• Storage temperature: 0°C to 50°C (32°F to 122°F)
• Humidity: 20% to 80%, noncondensing
• Altitude: Operation up to 2400 m (7874 ft.) above mean sea level
• Physical dimensions: 43 cm (17 inches) W x 50.5 cm (20 inches)
D x 24.5 cm (9.5 inches) H
• Weight: maximum of 25 kg (60 lbs.)
• UL installation category: UL Installation Category II, as defined
in Annex J of UL 61010A-1
• Pollution degree: UL Pollution Degree 2, as defined in Section
3.7.3.2 of UL 61010A-1
• Shipping and storage: The allowable shipping and storage
temperature and humidity ranges are 0°C to +50°C and 20-80%
noncondensing, respectively
• Input voltage range: 100 - 240 Vac, 1.5 Amp, 47-63 Hz or 100 -
120 Vac, 1.4 Amp, 47-63 Hz. (refer to the serial number label on
the rear panel of the instrument)
• AC inlet fuse: 3 Amp, 250 Vac, fast acting or 2 Amp, 250 Vac,
fast acting (refer to the fuse label on the rear panel of the
instrument)
Optics • Reporter laser: 532 nm, nominal output 10-16.5 mW, maximum
500 mW, frequency-doubled diode; mode of operation,
continuous wave (CW)
• Classification laser: 635 nm, 8 mW, maximum output 25
mW, diode; mode of operation, continuous wave (CW)
• Reporter detector: Photomultiplier tube, detection bandwidth of
565 - 585 nm
• Classification detector: Avalanche photo diodes with temperature
compensation
• Doublet discrimination detector: Avalanche photo diodes with
temperature compensation
Fluidics • Sheath flow rate: 90 µL ±5 µL/seconds
• Cuvette: 200 micron square flow channel
• Sample injection rate: 1 µL/second
• Sample uptake volume: 20 - 200 µL