Run Self
Diagnostics
Run self diagnostics to see if the device and all system operations are
running correctly. The system tests these functions when running self
diagnostics:
Flash memory test
Microcontroller RAM test
Nonvolatile memory test
DSP program CRC test
DSP capture test
High voltage module
Channel backgrounds test
Pressurization test
Actuator test
Syringe pump test
Backflush valve test
To run self diagnostics:
1. Click the Self-Diagnostics button from the maintenance tab. A
confirmation dialog box appears.
2. Click OK. The system begins processing the various self
diagnostic tests. As it completes the tests, the status bar changes
from a “Processing” command state to an “Idle” command state.
The self-diagnostic test should take less than one minute to
complete. If the self diagnosis failed, the system lets you obtain
more detailed information regarding the results of the self-
diagnostic test.