User Manual, Version 2.1

Luminex 100 IS User Manual Version 2.1 xMAP Technology
5 - 4 PN 89-00002-00-061 Rev. A
Accuracy and Precision Sample uptake volume ± 5%
Classification of microspheres > 80%
Misclassification of microspheres < 0.5%
Temperature control 0°C to +2°C of target
Internal sample carry over < 0.9%
Soluble background fluorescence emission at 575 nm
automatically subtracted from fluorescence intensity values
Sensitivity Detect 1000 fluorochromes phycoerythrin (PE) per microsphere
Reporter channel dynamic range: 3.5 decades
Capacity The specifications below reflect minimum capacity values:
10 GB hard drive
Store data for up to 240,000 test results
1.4 MB 3½” floppy disk
100 MB read/write CD ROM
Analyze multiple 96-well plates per batch
Analyze multiple assay templates per plate
Distinguish a minimum of 1 to a maximum of 100 unique xMAP
microsphere sets in a single sample
Detect and distinguish surface reporter fluorescence emissions at
575 nm on the surface of 1-100 unique xMAP microspheres sets
in a single sample
Reporter channel range: 3.5 decades of detection
Sample core: 15 - 20 µm core at 1 µL/sec sample inject rate
Maintain samples at a constant temperature from 35°C to 55°C
(95°F to 131°F)
Automatic sampling from a 96-well plate
Start sampling from any well position
Sheath container and waste container hold enough volume to run
up to two 96-well plates between refills
Microtiter plates with 96 wells must be compatible with the XYP
instrument plate holder. The following microtiter plate types are
compatible with the XYP instrument plate holder: flat bottom,
conical, round, filter bottom, half plates [overall height no more
than 0.75” (19 mm)], any color
Microtiter plates with 96 wells must be compatible with XYP
instrument heater block temperature from 35°C to 55°C (95°F to
131°F) when performing heated assays and using the heater
block.