User Manual, Version 2.1

xMAP Technology Troubleshooting
PN 89-00002-00-061 Rev. A 9 - 9
Sample Empty Detected This error message occurs when the system completely acquires the
entire sample volume.
See if the samples were properly mixed before they were
dispensed into the wells.
See if you loaded the sample to the correct well on the plate.
Check if the sample volume added to the well is correct.
Remove and sonicate or flush the XYP instrument Sample Probe
to remove possible clogs.
Check if the XYP instrument Sample Probe is properly aligned.
Complete the following analyzer functions from the Maintenance
tab: Drain, Backflush, and Alcohol Flush.
Run the system controls.
Check to see if the delta cal temperature <
3.0 degrees.
Recalibrate if greater than 3.0 degrees and retry.
If the system controls pass, contact the kit manufacturer.
Unknown Formula Failure Ensure that you selected the correct template.
Verify that the system is functioning properly by running system
controls.
If the system controls pass, contact the kit manufacturer.
Luminex 100
Analyzer Problems
Power On Indicator Light
Is Not On
Ensure that the Luminex 100 analyzer is powered on.
Ensure that the system is plugged into an outlet with current.
If using a surge protector, see if the power strip is On.
Remove and reconnect the cable.
Check the fuses (see “Fuses” on page 8-10).
PC Is Not Powered On Ensure that the PC voltage is set correctly by checking the back
of the PC. The options are 115 Volts and 230 Volts.
Ensure that the system is plugged into an outlet with current.
If using a surge protector, see if the power strip is On.
Remove the cable and reconnect.
Power on the PC.
Sheath Fluid Is Leaking
Onto the Plate Holder
Remove and sonicate or flush the XYP instrument Sample
Probe. Load a new plate onto the plate holder.