User Manual, Version 2.1
Luminex 100 IS Maintenance Log Month(s): ________________________________________Year:_______________
Maintenance Log
Procedure: To use this log, fill in the month(s) and year above. This form is designed to record information for four weeks. Fill in the dates for which this log
will be used in the first line of the table below. For each item listed at the left, fill in your initials under each date that you perform the item.
Note: Follow your standard laboratory safety practices when cleaning or maintaining the system. Do not remove the analyzer cover under any circumstances.
Dates
Daily maintenance
Startup Initials: (for each item listed at the left, fill in your initials under each date that you perform the item)
Laser warmup
Check sheath fluid
Check waste level
Tighten sheath cap
Alcohol Flush (70% isopropanol)
Wash twice (sheath fluid)
Shut down Initials: (for each item listed at the left, fill in your initials under each date that you perform the item)
Sanitize (20% household bleach
solution)
Wash twice (water)
Loosen sheath cap
Turn system off (optional)
Weekly
Visual inspection Date/Initials: Date/Initials: Date/Initials: Date/Initials:
Monthly
Clean Sample Probe Date/Initials:
Wipe exterior surfaces Date/Initials:
Calibrate and verify Date/Initials:
Every six months
Replace air intake filter,
analyzer
Date/Initials:
Replace air intake filter, XYP Date/Initials:
Replace syringe plunger seal, or
syringe
Date/Initials:
Check analyzer ventilation filter Date/Initials:
Yearly
Replace sheath filter Date/Initials:
As required
Replace fuses Date/Initials:
Comments:
PN 89-00002-00-061 Rev. A