Manual
Luminex 100 IS User Manual Version 2.3 xMAP Technology
5 - 26 PN 89-00002-00-072 Rev. C
Figure 5-18. Non-Competitive Assay
Examples of out-of-range sample labeling for competitive assays are
shown in Table 5-7. Figure 5-19 shows what a standard curve for a
competitive assay should look like.
Table 5-6. Non-Competitive Out-Of-Range Labeling
Condition Concentration Label MFI of Standard Referenced
Left of Curve < min concentration standard Lowest
Right of Curve > max concentration standard Highest
Table 5-7. Competitive Out-Of-Range Labeling
Condition Concentration Label MFI of Standard Referenced
Left of Curve < min concentration standard Highest
Right of Curve > max concentration standard Lowest