Figure 43. The System Monitor
The system can diagnose real-time system problems related to fluid
input and output. The system can also detect and report if the
Luminex XYP heater block temperature is out of range or
unacceptable temperature conditions, including these items:
Luminex XYP heater block temperature time-out
heat circuit failure
temperature out-of-range sensing
temperature change since calibration, or a change in channel
temperature
Table 4 defines the values listed in the System Monitor. These values
are useful for diagnostic purposes when communicating with
Luminex Customer Support.
Table 4. System Monitor Values
Property Value Units of Measure
Air Pressure PSI, air pressure to the sheath
container from the air pump
Sheath Pressure PSI, Sheath pressure from the
sheath container through the
system
Delta Cal Temperature °C, temperature deviation from
the last calibration
Board Temperature °C, temperature of the analog
board
DD Temperature °C, DD temperature at the U
block inside the optics platform
CL1 Temperature °C, CL1 temperature at the U
block inside the optics platform