the command process in yellow text below the graphic. For example,
Figure 36 shows the Warmup command graphic.
Figure 36. Processing Device Activity Commands
Table 3 shows a recommended schedule for maintenance operations.
Table 3. Maintenance Operations:
Recommended Use Schedule
Operation Recommended Use Schedule
Warmup Daily
After four hours of system inactivity
Prime Daily
To remove air from sheath fluid tubing
After performing these actions:
– refilling the sheath container
– removing and replacing sheath container
– changing the sheath fluid filter
– changing the syringe seal
Backflush
Troubleshooting purposes only:
– to remove obstructions from the cuvette
– if fluid does not flow through the waste
tubing during prime cycles or during
sample acquisition
– if fluid drips from the Sample Probe during
priming and forms puddles of fluid on the
plate
Alcohol Flush Daily
After changing the Sample Probe
To remove air bubbles from the cuvette using
70% isopropanol