xMAP Technology Using Luminex 100 IS 2.2 Software
PN 89-00002-00-069 Rev. A 6 - 25
the command process in yellow text below the graphic. For example,
Figure 36 shows the Warmup command graphic.
Figure 36. Processing Device Activity Commands
Table 3 shows a recommended schedule for maintenance operations.
Table 3. Maintenance Operations:
Recommended Use Schedule
Operation Recommended Use Schedule
Warmup • Daily
• After four hours of system inactivity
Prime • Daily
• To remove air from sheath fluid tubing
• After performing these actions:
– refilling the sheath container
– removing and replacing sheath container
– changing the sheath fluid filter
– changing the syringe seal
Backflush
• Troubleshooting purposes only:
– to remove obstructions from the cuvette
– if fluid does not flow through the waste
tubing during prime cycles or during
sample acquisition
– if fluid drips from the Sample Probe during
priming and forms puddles of fluid on the
plate
Alcohol Flush • Daily
• After changing the Sample Probe
• To remove air bubbles from the cuvette using
70% isopropanol