Manual
Luminex 100 IS User Manual Version 2.2 xMAP Technology
6 - 12 PN 89-00002-00-069 Rev. A
The Status Bar dynamically displays status information as the
software processes commands. Errors appear in the Message Log at
the bottom of the Diagnostics tab.
Text on the Diagnostics tab turns red if the system encounters an
error. The Message Log on the Diagnostics tab indicates where the
error occurred.
• Idling: waiting for a command
• Aspirating: drawing in sample
• Collecting Data: collecting data
• Sanitizing: sanitizing the instrument
• Washing: washing the instrument
• Priming: priming the instrument
• Calibrating: calibrating the instrument
• Canceling: canceling a command
• Backflushing: backflushing the instrument
• Draining: draining the instrument
• Warming Up: warming up the instrument
• Verifying: verifying calibration
• Soaking: soaking the probe
• Adjusting Sample Probe
• Self Diagnosing: performing self-diagnostic
routine
Laser Status Laser temperature and readiness status
Green Lasers warmed up
Yellow Warmup timer countdown in seconds from 1800
Red Lasers off
Total Events/
Second
Number of total bead events detected per second
Region
Events/
Second
Number of sample reactions detected per second
Table 1. Status Bar Color Coding (Continued)
Category Color Indicates