Performance
Specifications
Speed Speed: greater than or equal to 933 MHz Pentium III processor
with 256 MB RAM
USB cabling for fast data transfer
Automatic transfer of assay templates and new reagent
information into the system using a 3½” inch floppy disk or large
capacity read/write CD
Installation < 4 hours
System calibration < 10 minutes
System controls < 10 minutes
Barcode reader entry of sample IDs
Automatic post analysis
Analyze one 96-well plate/hour depending on manufacturer’s kit
Up to 100 xMAP microsphere sets per sample
Sheath flowrate 90 µL/sec. ± 5 µL
Sample injection rate into detector area 1 µL/sec. ± 0.05 µL
System warmup 30 minutes: Systems that remain inactive for at
least four hours will require a warm-up to restart the lasers. After
acquiring sample, running system calibrators, running system
controls, and warming up the instrument, the system resets the
four-hour internal clock.
Accuracy and Precision Sample uptake volume accuracy ± 5%
Classification of xMAP microspheres > 80.0%
Misclassification of xMAP microspheres < 0.5%
Temperature control 0°C to ± 2°C of target
Internal sample carry over < 0.9%
Soluble background fluorescence emission at 575 nm
automatically subtracted from fluorescence intensity value
Sensitivity Detect 1000 fluorochromes Phycoerythrin (PE) per microsphere
Reporter channel dynamic range: 3.5 decades