Installation Guide Instruction Manual

xMAP Technology Luminex 100 IS System Specifications
PN 89-00002-00-028 Rev. B 7
Uninterruptible
Power Supply
(UPS)
Luminex highly recommends using an uninterruptible power supply
(UPS) to protect your system from power outages. Choose a supply
that can provide 1050 Watts for at least 45 minutes. The UPS should
be UL listed, CSA certified, and CE marked when used
internationally.
Surge Protector If you do not use a UPS, use a surge protector. Choose a protector
that meets your needs. Factors to consider include electrical
environment, endurance, suppressed voltage rating, and method of
protection. It should have six outlets, rated at least 1500 Watts, and
be UL listed, CSA certified, CE marked for nondomestic use when
used internationally.
Luminex 100 IS
System
Specifications
Luminex 100 Analyzer Indoor use only
Operating temperature: 15°C to 30°C (59°F to 86°F)
Humidity: 20% to 80%, noncondensing
Altitude: Operation up to 2400 m (7874 ft.) above mean sea level
Physical dimensions: 43 cm (17 inches) W x 50.5 cm (20 inches)
D x 24.5 cm (9.5 inches) H
Weight: maximum of 25 kg (60 lbs.)
UL installation category: UL Installation Category II, as defined
in Annex J of UL 61010A-1
Pollution degree: UL Pollution Degree 2, as defined in Section
3.7.3.2 of UL 61010A-1
Shipping and storage: The allowable shipping and storage
temperature and humidity ranges are 0°C to + 50°C and 20-80%
noncondensing, respectively
Input voltage range: 100 - 120 V~ ±10%, 1.4 Amp, and 200-240
V~ ±10%, 0.8 Amp, 47-63 Hz.
AC inlet fuse: 3 Amp, 250 V~, fast acting
Luminex XYP Instrument Ambient temperature: 15°C to 30°C (59°F to 86°F)
Humidity: 20% to 80%, noncondensing
Altitude: operation up to 2400 m (7874 ft) above mean sea level
Physical dimensions: 44 cm (17.25 inches) W x 60 cm (23.5
inches) D x 8 cm (3 inches) H