Installation Guide Owner's manual
Luminex 100 IS 2.1 User Manual Installation Guide xMAP Technology
8 PN 89-00002-00-105 Rev. A
• Weight: 15 kg (33 lbs.)
• UL installation category: UL Installation Category II, as defined
in Annex J of UL 61010A-1
• Pollution degree: UL Pollution Degree 2, as defined in Section
3.7.3.2 of UL 61010A-1
• Heater operating range: 35°C to 55°C (95°F to 131°F) with
tolerance 0°C to +2°C
• Luminex XYP instrument, input voltage range: 100-240 V~
±10%, 1.8 Amps, 47-63 Hz
• AC inlet fuse: 3 A, 250 V~, fast acting
Luminex SD System • Ambient temperature: 15°C to 30°C (59° to 86°F)
• Humidity: 20% to 80%, noncondensing
• Altitude: designed to operate at up to 2400m (7874 feet) above
mean sea level
• Physical dimensions: 20 cm (8 inches) W x 30 cm (11.75 inches)
D x 24.75 cm (9.75 inches) H
• Weight: 9 kg (20 lbs)
• UL installation category: UL Installation Category II, as defined
in Annex J of UL 61010A-1
• Pollution degree: UL Pollution Degree 2, as defined in Section
3.7.3.2 of UL 61010A-1
• Input voltage range: 100-240 V~ ±10%, 0.4 Amps, 47-63 Hz
• AC inlet fuse: 2 Amp, 250 V~, time lag
PC Specifications • CE marked and UL listed
• Power cords: Power cords specific to the country of use are
included in the system as required
• Power: 115-230 V~, 6 Amps, 50-60 Hz
Hardware Components The Luminex 100 IS system includes the following hardware
components:
• Luminex 100 analyzer
• Computer (PC), monitor, and accessories
• Luminex XYP instrument
• Luminex Sheath Delivery System (Luminex SD™) (optional)
• Power cables
• Alignment guide
• One short sample probe
• Luminex XYP instrument sample probe