Low Air Loss Therapy Mattress Replacement System Model # 763400A Operators Manual Please read the manual before use GF0500067REVA05
Contents Important Safeguards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
IMPORTANT SAFEGUARDS READ ALL INSTRUCTIONS BEFORE USING DANGER - To reduce the risk of electrocution: 1. Always unplug this product immediately after using. 2. Do not use while bathing. 3. Do not place or store product where it can fall or be pulled into a tub or sink. 4. Do not place in or drop into water or other liquid. 5. Do not reach for a product that has fallen into water. Unplug immediately. WARNING - To reduce the risk of burns, electrocution, fire, or injury to persons: 1.
1. Introduction This manual should be used for initial set up of the system and for reference purposes. 1.1 General Information LumexAir systems use most advanced technologies and mattress materials to deliver the best quality, comfort and whisper-quiet operations to you and your customers. Most systems in LumexAir are recommended for prevention and treatment of pressure ulcers up to and including stage IV pressure ulcers.
2. Product Description 2.1 Pump and Mattress System The system consists of a high volume airflow blower and numerous large air loss holes on the mattress to achieve the true low air loss therapy. It also provides excellent ventilation and keeps patients dry from humidity. The 10 inch high mattress, which consists of 8 inch high air bladders at the top and a layer of 2" foam at the bottom, provide maximum comfort for patients.
3. Installation 3.1 Unpacking The pump unit and mattress are packaged in a separate box to secure its contents inside. Unpacking these boxes to remove the pump unit and mattress and check for any damage, which may have occurred in shipping. If there are damages, please contact your dealer immediately. 3.2 Setting Up 1. Place the mattress on top of the bed frame. Please note for the foot end. Mattress can be secured firmly by fixing the elastic straps at the bottom bed frame.
4. Plug into main electrical outlet. NOTE 1. Make sure the pump unit is suitable for the local power voltage. 2. The plug also serves as the disconnect device CAUTION: The pump can only be applied to the mattress recommended by the manufacturer. Do not use it for any other purpose. 5 Then turn the main power switch ON. The power switch is located on the right side panel of the pump unit. NOTE: Ensure the CPR connectors are sealed completely.
4. Operation NOTE: Always read the operating instructions before use. 4.1 Function Description: Control Button Function Description POWER Two conditions: 1. To turn on control unit. It takes approximately 1 minute to inflate the mattress. The default setting is at a comfort range, level 6, for the quickest inflation time. 2. To turn off Power Failure Alarm in the event of a sudden power outage during normal use. ALARM MUTE Enable and disable the audible alarm feature.
Control Button Function Description AUTOFIRM For maximum inflation of mattress. All other functions are disabled unless this button is pressed again. This mode will last for 20 minutes and then it will return to default setting automatically. STATIC For true low air-loss therapy mode, similar to a regular air mattress with very low air pressure support and true low air-loss therapy.
4.2 Operating Instruction 1. Press the POWER button on the display panel to start the system. 2. For faster inflation, the AUTOFIRM function can be activated. This mode will last for 20 minutes and then it will return to default setting automatically. 3. After the mattress is fully inflated, user can change the setting according to their preference or recommended by physician. 4.3 CPR When there is an emergency to perform CPR on the patient, quickly pull the CPR valves to release air from mattress.
NOTE: If the pressure level is consistently low, check for any leakage at air cells or connecting tubes. If necessary, replace any damaged cells or tubes or contact local qualified dealer for repair. If the control unit is equipped without audible alarm feature, the Low Pressure Indicator will flash until the low-pressure fault condition is solved.
1. 2. 3. 4. 5. 6. Storage Pull the CPR tags and make sure all cells are deflated. Lay the mattress out flat and upside down. Protect the air tube coupler by putting it in the inner side of mattress. Roll the mattress from the head end towards the foot end. Foot-end straps can then be stretched around the rolled mattress to prevent unrolling. 7. Maintenance 7.1 General 1. Check main power cord and plug if there is abrasions or excessive wear. 2. Check mattress' cover for signs of wear or damage.
8. Troubleshooting Problem Power is not ON Alarm is on (audible & visual) Solution • Check if the plug is connected to main power. • Check if there is any blown fuse. • Check if the connection between air tube connector to pump unit is tightly secured. • Check if the CPR is sealed. • Check if there are any leakage on air cells. • Check if all tubing connections along mattress are secured.
9. Technical Description Specifications: Pump Specification Power Supply 120V AC 60Hz, 1A (For 120V system) Fuse Rating 5A (For 120V system), T3A ( For 230V system) (Note: See rating label on the product) Dimension 37 x 27 x 13 (cm) or 14.4" x 10.5" x 5" Weight 4.5 kg or 9.