Owner manual

M345X12 Blackbody Manual Operation 12
3.2.1 RS485 Option with RS-485 to RS-232 Converter
RS-485 Two wire is available as an option. A typical connection scheme with a converter is shown
below:
Figure 5: RS-485 for multi-drop applications
3.2.2 Changing the Controller Communication Address
The controller has a distinct address. The factory address setting is 1.
To change the address:
1. Press PAGE button. The word ACCESS will appear in lower display.
2. Press SCROLL button. LEVEL 1 will appear.
3. Press UP button until LEVEL 3 appears.
4. Enter PASSCODE by Pressing UP button until ‘3’ appears and wait until PASS’ appears.
5. Press PAGE button repeatedly until ‘COMMS’ ‘H module’ appears.
6. Press UP button until J module’ appears.
7. Press SCROLL button twice until ‘ADDRESS’ appears.
8. Press UP/DOWN buttons to change address.
9. Press PAGE button until ‘ACCESS’ appears.
10. Press SCROLL button. Display will show ‘LEVEL 3’.
11. Press DOWN button until ‘LEVEL 1’ appears.