2.2 Technical Data
This Model Model Temperature Range
[ ] M390-S 600 to 3000 °C (1110 to 5430 °F) Custom split design
Accuracy:
±0.25% of reading ±1 C
Cavity:
Closed end graphite
tube 150
mm (6.0") long
with 64 mm (2.50")
heated length
Source Diameter:
25
mm (1.00")
Emissivity
:
0.995
±
0.003
Cooling
:
Water cooled. Standard hose c
onnection on rear, 3 lpm (1 gpm)
Cooling Inlet Pressure:
90 PSI maximum
Customer is responsible to provide pressure safety valves at inlet
and outlet. Outlet (drain) must not be blocked.
Purge Gas:
Argon High Purity (99.99
% pure)
Purge Gas Inlet Pressure:
25 PSI maximum.
Customer is responsible to provide over
pressure safety valves at inlet
Heating Element Type:
Gr
aphite tube argon gas purge, 80
lph (4
-
8
cfh)
Temperature Sensor:
Mikron "Infraducer"
Method of Control:
Digital PID Controller
Remote Set Point:
RS
422 serial communication output
Warm
-
up Time:
5
minutes from ambient to 2300 °C
Operating Ambient Temp.:
0 C to 44 C (32 F to 110 F)
Operating Ambient
Humidity:
0 to 90% RH non
-
condensing
Power Requirements:
[ ] 208 or [ ] 220 or [ ] 240 VAC ±10% [ ]50 [ ]60Hz
[ ] 15 kVA [ ] 24 kVA (specified at time of order)
Power
Connector:
Customer supplied.
[ ] 15kVA Must use plug that complies with IEC-309 at 75 Amps at 220VAC
[ ] 24kVA Must use plug that complies with IEC-309 at125 Amps at 220VAC
Dimensions:
Blackbody Enclosure:
Controller:
34.5” (87.6 cm) H x 22” (55.9 cm) W x 38.5” (97.8 cm) D
19” Rack mount, 7” (17.8 cm) H x 18” (46 cm) D.
Maximum Weight:
182 kgs (400 lbs)