Manual

M330 Manual Introduction 9
2 Introduction
2.1 Description
The M330 is a blackbody calibration source utilizing a digital indicating temperature controller
that may be set to any temperature between 300 °C (572 °F) and 1700 °C (3092 °F). A precision
thermocouple is embedded in the blackbody cavity providing high accuracy and repeatability.
The temperature controller uses the industry standard PID algorithms to control the emitter
temperature to within +/-0.5 °C. The blackbody uses a resistive heater that provide long life.
Running at temperatures greater than 1600 °C for long periods will shorten heater life.
2.2 Technical Data
Temperature Range:
300 C to 1700 C or 572 F to 3092 F
Accuracy:
±0.25% of reading ±1 C
Temperature Resolution:
0.1 C
Stability:
0.5 C per 8 hour period
Source Non
-
Uniformity:
Non-Uniformity: ±1 C within center 1/3 of ID.
Cavity:
Closed end tube 41
mm ID X 255
mm long with 125
mm heated
length.
Aperture Diameter:
1.00 (25 mm)
Effective
Emissivity
1
.0
a
t
1
-
1.7
µ
m
and 0.65
µ
m
(
radiance calibration only
)
Method of Control:
Digital Self
-
Tuning PID Controller
Warm
-
up Time:
45 min
utes
from ambient to 300
°C.
65 to 80 minutes from 300 to 1600 C
Operating Ambient Temp.:
0 C to 44 C (32 F to 110 F)
Cooling
:
Fan cooled, air
inlet on rear panel
Power Requirements:
[ ]208 [ ]220 [ ]230 [ ]240
VAC ±10% 50/60
Hz 3000
VA
Maximum Weight:
175 lbs. (80 kgs)
2.3 Calibration Certificate
The certificate of calibration of your blackbody source has been enclosed as a part of this
manual or under separate cover. This certificate contains valuable information including
traceability number to NIST. Please observe recommended date of next calibration.