Confidential Information The material contained herein consists of information that is the property of LumaSense Technologies and intended solely for use by the purchaser of the equipment described in this manual. All specifications are subject to change without notice. Changes are made periodically to the information in this publication, and these changes will be incorporated in new editions.
Contents 1 General Information...................................................................................................... 5 1.1 Information about the user manual..................................................................... 5 1.1.1 Legend ................................................................................................................. 5 1.2 Safety ..................................................................................................................... 5 1.2.
To ensure consistent document formatting, this page was intentionally left blank.
1 General Information 1.1 Information about the user manual Congratulations on choosing the high quality and highly efficient Mikron blackbody. This manual provides important information about the instrument and can be used as a work of reference for installing, operating, and maintaining your blackbody. It is important that you carefully read the information contained in this manual and follow all safety procedures before you install or operate the instrument.
warranty information. This warranty covers manufacturing defects and faults which arise during operation, only if they are the result of defects caused by LumaSense Technologies. The warranty is VOID if the instrument is disassembled, tampered with, altered, or otherwise damaged without prior written consent from LumaSense Technologies; or if considered by LumaSense Technologies to be abused or used in abnormal conditions. 1.
The instrument must be shipped in the original packing container or its equivalent. LumaSense Technologies is not responsible for freight damage to instruments that are improperly packed. Contact us to obtain an RMA number (if one has not already been assigned by Technical Support). Clearly indicate the assigned RMA number on the shipping package exterior. Send RMA Shipments to your nearest technical service center: Santa Clara, California Frankfurt, Germany LumaSense Technologies, Inc.
To ensure consistent document formatting, this page was intentionally left blank.
2 Introduction 2.1 Description 2.2 Technical Data The blackbody calibration source features a unique uniformly heated spherical cavity with a near ideal emissivity of 0.999 or better. The M300 will deliver any temperature between 200 °C and 1150 °C. A convenient microprocessor based, self-tuning digital PID controller holds the selected temperature to within 1 °C, assuring high accuracy calibration of infrared thermometers, thermal imaging equipment, etc.
2.
3 Installation and Operation 3.1 Installation Install the blackbody emitter in the desired location that preferably has no moving air near the emitter aperture. The M300 has been factory wired to operate from your standard 208/220 or 240VAC 20 ampere supply line as per your order. 3.2 Operation 3.2.1 Starting Up 1. Allow the M300 to warm-up to room temperature for at least two hours after unpacking and prior to applying power to remove any condensation. 2.
BLACKBODY TEMPERATURE BLACKBODY SETPOINT Eurotherm controller 3.2.3 Using Temperature Controller Tuning Constants PID Set 1 (T<337 °C) PID Set 2 (T>337 °C) PB1 ________ PB2 _________ INTEGRAL _________seconds INT2 ___________seconds DERIVATIVE________seconds DERV2 _________seconds 3.2.
6. Press UP button until J module appears. 7. Press SCROLL button twice until ADDRESS appears. 8. Press UP/DOWN buttons to change address. 9. Press PAGE button until ACCESS appears. 10. Press SCROLL button. Display will show LEVEL 3. 11. Press DOWN button until LEVEL 1 appears. 12. Press the center, round button on top row to go back to the normal display mode. A CD containing a sample program is included with your order.
To ensure consistent document formatting, this page was intentionally left blank.
4 Maintenance 4.1 Cavity Life The life expectancy of the cavity depends on the time and temperature of operation. The higher the temperature, the shorter the life span. At temperatures near the upper end of the range, the life expectancy is shortened by approximately 50%. We recommend an occasional visual check be made to insure maximum system integrity.
To ensure consistent document formatting, this page was intentionally left blank.
Appendix A: Declaration of Conformance M300 Manual Appendix A: Declaration of Conformance 17
To ensure consistent document formatting, this page was intentionally left blank.
Index A Address 12 Appropriate use 9 C Cavity Life 15 Cleaning 15 D Dimensions 10 Disposal 7 F Fuse Locations 15 G General Information 5 I Installation 11 L S Safety 5 Safety Interlocks 15 Serial Communication 12 Service Request 6 Setpoint 11 Shutting Down 13 Starting Up 11 Storage 7 Support 6 T Technical Data 9 Temperature Controller Tuning Constants 12 Transport 7 U Unpacking the Instrument 6 W Warranty 5 Legend 5 Liability 5 M Maintenance 15 O Operation 11 P Packing 7 Periodic Inspections 1