Manual

IS 50-LO plus / IGA 50-LO plus Manual Introduction 11
Storage temperature:
-
20 …
70
°C
Protection class:
IP5
4
Weight:
Converter: appro
x. 600 g
Optical head II: approx. 140 g
Fiber (2.5 m): approx. 250 g
Dimensions:
125 mm x 80 mm x 57 mm (L x B x H)
Dimensions (Optics):
See Section 2.4
Mounting position:
Any
CE label:
According to EU directives about electromagnetic immunity
*Additional measurement uncertainty due to offset drift of the signal converter at long
temperature ranges. Tc = measuring temperature up to which an additional measurement
uncertainty occurs when the ambient temperature differs from the reference temperature of
23 °C
Type
MB/
°C
Tc/
°C
IS 50
-
LO plus
, MB
25
750…2500
760
IGA
50
-
LO
plus
, MB 13
300…1300
335
IGA
50
-
LO plus
,
MB 13
.
5L
250…1350
343
IGA 50-LO
plus
, MB 18
350…1800
400
IGA 50-LO
plus
, MB 25
450…2500
462
IGA 50-LO
plus
, MB 20
300…2000
420
IGA 50-LO
plus
, MB 25
350…2500
362
Note: The calibration / adjustment of the instruments was carried out in accordance
with VDI/VDE directive “Temperature measurement in industry, Radiation
thermometry, Calibration of radiation thermometers”, VDI/VDE 3511, Part 4.4.
For additional details on this directive, see http://info.lumasenseinc.com/calibration or
order the directive from “Beuth Verlag GmbH” in D-10772 Berlin, Germany.
2.4 Dimensions
Converter:
Optical head I:
Optical head II:
(fixed adjusted)
Optical head II:
(focusable)
All dimensions in mm