Instruction Manual
IS 12 AI & IS 12-AI/S Data format UPP (Universal Pyrometer Protocol) 33
9 Data format UPP (Universal Pyrometer
Protocol)
Via interface and a suitable communication software or via “Test” function of the InfraWin
commands can be exchanged directly with the pyrometer.
The data exchange occurs in ASCII format with the following transmission parameters:
The data format is: 8 data bits, 1 stop bit, even parity (8,1,e).
The device responds to the entry of a command with: output (e.g. the measuring value) + CR
(Carriage Return, ASCII 13), to pure entry commands with “ok” + CR.
Every command starts with the 2-digit device address AA (e.g. “00”). This is followed by 2 small
command letters (e.g. “em” for level of emissivity e), finished with CR
This is followed, if necessary for that command, by the ASCII parameter “X”. If this parameter
“X” is omitted, then the device resets with the current parameter.
A „?“ after the small command letters answers with the respective settings (only at setting
commands, not at enquiry commands).
Example: Entry: “00em“ + CR
The emissivity setting (e) of the device with the address 00 is returned
Answer: “0970“ + CR means Emissivity = 0.97 or 97.0%
Description
Command
Par
ameters
Reading temperature
value:
AAms
Output: XXXXX (dec., in
1
/
10
°C or °F)
last digit is the decimal place (88880 = Temp.-Overflow)
Reading temperature
value repeated:
AAmsXXX XXX = 000…999 (XXX = number of measuring values)
Reading basic
temperature range:
AAmb
Output: XXXXYYYY (hex 8
-
digit, °C or °F)
XXXX = beginning of temp. range
YYYY= end of temp. range
Reading temperature sub
range:
AAme same as mb
Reading emissivity:
AAem
Output: XXXX (dez. 0010 … 1000 in ‰)
Emissivity:
AAemXX
XX = (10…99%), 00 = 100% (decimal)
Emissivity:
AAemXXXX
XXXX = (0010 … 1000‰) (decimal)
Exposition time t
90
: AAezX
X = 0 … 6 (decimal)
0 = intrinsic time constant
1 = 0.01 s 3 = 0.25 s 5 = 3.00 s
2 = 0.05 s 4 = 1.00 s 6 = 10.00 s
External deletion:
AAlx
Simulation of an external deletion contact
Delete maximum value: AAlzX
X = 0 … 8 (dez.)
0 = Maximum value storage off
1 = 0.01 s 4 = 1.00 s 7 = external deletion
2 = 0.05 s 5 = 5.00 s 8 = automatically deletion
3 = 0.25 s 6 = 25.00 s
Setting of
temperature
sub range:
AAm1XXXXYYYY
XXXX (hex 4
-
digit) beginning of temp. range (°C)
YYYY (hex 4-digit) end of temp. range (°C)
Analog output:
AAasX
X = 0 or 1; 0 = 0 to 20 mA
1 = 4 to 20 mA