Owner manual

Chapter 5
_________________________________________________________________________________________
_____________________________________________________________________________
BE6021-11 1316A-2 Multi Gas Monitor LumaSense Technologies A/S
Instruction Manual Page 90 of 97
3. Remove the fuses from the holders and check that they have
the correct rating (T 1.6 A slow-blow), see figure 5.3.
4. If the installed fuses do not have the correct rating or are
faulty, install the correct fuses.
5. Screw the fuse holders into the monitor.
6. Attach the mains cable.
5.1.3 Checking the Pump
Pull down the Bench Commands menu.
Select Pump On/Off
Following window will appear.