2. Type in the correct concentrations for the ticked CO2, Freon
gas.
The gas concentration must be the same as that stated on the
calibration gas bottle.
3. Connect the calibration-gas bottle as described in section
4.3.4.
4. Open the gas valve and adjust the flow so that a small surplus
flow exists in the side-stream tube.
5. Press the Send button to start the gain calibration.
After making the Span Calibration please make sure that the Mode is
set back into 5 and that the Zero calibration is set to “Auto”, as
described in section 4.3.5.
The monitor is now ready for making measurements.
Press Send.
The Process State In Progress is displayed in the Span Calibration
window (see figure 4.30) and the Data Status in the Status window is
changed to show Freon Span, see figure 4.31.
The 1316A-2 is calibrated with the selected gas at the concentration
keyed in.
Figure 4.30. Span Calibration window showing Freon calibration in
Progress.