Owner manual

Chapter 4
_________________________________________________________________________________________
_____________________________________________________________________________
BE6021-11 1316A-2 Multi Gas Monitor LumaSense Technologies A/S
Instruction Manual Page 74 of 97
Tick the gas(ses) to be Span calibrated, and its concentration value(s), see
figure 4.23. Press the “Send” button to perform the Span calibration. Observe
the “Process State” change to “In Progress” during the Span calibration, see
figure 4.24.
Figure 4.23 The Span Calibration Window to select the gasses and the
concentrations for Span calibration
The text box in the Span Calibration window now displays Span Calibration ACK.
The Span Calibration is in progress.