BE6021-11 1316A-2 Multi Gas Monitor LumaSense Technologies A/S
Instruction Manual Page 70 of 97
Figure 4.16 Main window displaying measurement started..
Open the status window and press the “Zero”-button in the Main window, in
order to perform a Zero calibration.
The ambient air is taken from the Zero reference inlet on the rear panel during
Zero calibration. In the Status window observe that the “Proc in prog”-status is
“In Progress”, while Zero calibration. is done, see figure 4.17.
When Zero calibration has successfully finished theProc in prog”-status is “Not
in Progress”, see figure 4.18.