BE6021-11 1316A-2 Multi Gas Monitor LumaSense Technologies A/S
Instruction Manual Page 58 of 97
Figure 4.3. The Status Window displaying Zero calibration in
Progress.
The Zero/O
2
Span is started controlled by the BZ6013 going through
the following steps:
i) The pump is turned OFF and all solenoid valves are deactivated.
ii) After a settling delay, the ambient pressure (Atm Press) is
measured and stored.
iii) The pump is turned ON.
iv) After a settling delay, the sample cell pressure (Abs Press) is
measured and stored.
v) The absolute pressure limit is checked (Abs Press Atm Press
Pressure Limit).
If this test fails, the Pump Out Flow Check is shown in the Status
window (see figure 4.4) and the zero calibration procedure is aborted: