BE6030-13 1314i Photoacoustic Gas Monitor LumaSense Technologies A/S
Page 18 of 199
If the baud rate displayed is incorrect press
and use and to
display the correct value. Press
again to store the selection.
If the baud rate displayed is correct, then press to continue to the
next parameter.
Press S1 to select 1 STOP BIT.
Press S1 to select 7 DATA BITS
Press S2 to select EVEN PARITY
Press S3 to select LEASED-LINE
Press S3 to select HARD-WIRED HANDSHAKE.
Press SET-UP to exit the set-up mode
Press RESET and S1 in order that the new settings are enabled.
The monitor and the Gas Monitoring Software 7304 are now able to
communicate together.
2.5.3 Checking/Changing the Text line Terminator.
The Text line Terminator must be set using the push-buttons on the
front of the monitor.
Press SET-UP S2 S3
Press SET-UP to exit the set-up mode
2.5.4 Checking/Changing the Data Log and Error Log
Options.
The Data Log and Error Log must be set using the push-buttons on
the front of the monitor.
Press SET-UP S3 S1 S1 S3 S1 S1
Press SET-UP to exit the set-up mode
2.6 Setting the Ethernet Communication Parameters
The LumaSoft Gas 7810 application with Ethernet communication is
able to communicate with the 1314i gas monitor through a local
Ethernet network.
The homepage of the 1314i gas monitor can also be displayed in a
web browser, when the 1314i gas monitor is connected to a local
Ethernet network.