Manual

Chapter 12
______________________________________________________________________
_____________________________________________________________________________
BE6030-13 1314i Photoacoustic Gas Monitor LumaSense Technologies A/S
Page 176 of 199
12.10.4 Performing a Span Calibration of the Water-vapour
Filter
Section 12.4.1 describes a good method of producing water-vapour
of known concentration. The directions for performing a span
calibration of water-vapour filter are similar to those for performing a
humidity-interference calibration of the other installed filters is as
described in section 12.7.2. In both calibration tasks the signal in the
cell is measured when a supply of zero gas containing a constant
concentration of water vapour is attached to the Monitor. There are,
however, two differences:
1. During a humidity-interference calibration the signal is measured
using each of the installed filters (including the water-vapour
filter), whereas during the span calibration of the water-vapour
filter the signal is only measured with the water-vapour filter.
2. During a humidity-interference calibration, it is not necessary to
know the concentration of water-vapour supplied to the cell as it
is only necessary to measure the size of each signal (with filters
“A” E”) relative to the size of the signal measured with the
water-vapour filter. However, during a span calibration of the
water-vapour filter it is necessary to know the exact
concentration of water-vapour supplied to the cell, because during
this calibration the conversion factor is calculated. This factor
relates the measured signal to the concentration of water-vapour
in the cell.
Follow the procedure described in section 12.7.2.
12.11 Setting-up a Combined Zero-point and
Humidity-interference Calibration Stand-
alone Use
We have explained in previous sections of this chapter (section 12.6
and section 12.7.1) that a zero-point calibration can only be
performed alone if a humidity gain factor is already stored in each
of the filter banks which are active during the zero-point calibration
task. Likewise, a humidity-interference calibration can only be
performed alone if a concentration offset factor is already stored
in each of the filter banks which is active during the humidity-
interference calibration task.
If the necessary calibration factors are not stored in the Monitor’s
memory, a zero-point calibration task has to be performed together
with a humidity-interference calibration task.