BE6030-13 1314i Photoacoustic Gas Monitor LumaSense Technologies A/S
Page 16 of 199
3. Use a screwdriver to remove the fuse-holder:
4. Remove the old fuse from the holder and replace it with a new.
Check that the fuse has the correct rating.
The fuse must be:
Slow-blow (T) fuse with a rating of 2.5 A, 250 VAC, 5x20 mm, UL
approved.
LumaSense order number: VF0102A
5. Install the fuse-holder and re-assemble the 1314i again.