Instruction Manual

Chapter 1
________________________________________________________________________
________________________________________________________________________
BE1285-15 1309 Multipoint Sampler LumaSense Technologies A/S
Page 9 of 42
1.3 Rear Panel
Fig.1.3 The rear panel of the 1309
AC Mains: A 3-pin connector accepting Power Cable for connection
to a single phase AC mains supply with protective Earth.
Mains Voltage: Connect 1309 to mains supply with 100-240 Vac, 50/60
Hz.
Outlet to Analyzer: Mounting stub for connecting the sampler system of the
1309 to the inlet of an INNOVA Gas Monitor via tubing.
See section 2.4.4. Usable models are 1302, 3425, 3426,
1312, 1314 or 1412. In the following referred to as “Gas
Monitor”. Please contact LumaSense Technologies A/S
for available models.
Waste Air Outlet: Mounting stub for connecting the external pump to the
1309’s sampler system. See section 2.4.4.
Interface IEEE-488: Digital interface designed in accordance with IEEE 488-
1978. The IEEE interface is identical in use to that de-
scribed in IEC 625-1; full compatibility is only a matter
of using the correct cables and connectors. For more de-
tail about the IEEE/IEC interface, refer to Chapter 4.
Device Address: Bank of 8 DIP switches which decide the 1309’s interface
address. See section 2.3.1 for details of how to set the
interface address.
00061-02-110311
50-60Hz 70VA
100 - 240 Vac
Fuses: T1 .6A L