Instruction Manual
Chapter 1
________________________________________________________________________
________________________________________________________________________
BE1285-15 1309 Multipoint Sampler LumaSense Technologies A/S
Page 8 of 42
1.2 Front Panel
Fig.1.2 The front panel of the 1309
Sampler: 12 mounting stubs for connection of tubing to sampling
points. Each stub is numbered, and has a correspond-
ingly-numbered lamp. When the lamp is lit, it indicates
that the corresponding sampling valve is open, see sec-
tion 3.3. The Analyzer/Waste Air lamps indicate which
way the internal 3 way valve is set, see section 3.3.
Temperature Sensors Input:
6 inputs suitable for use with the INNOVA Air Tempera-
ture Transducer type MM0034, Surface Temperature
Transducer MM0035 or Operative Temperature Trans-
ducer type MM0060.
Interface: 3 lamps which indicate the function of the IEEE inter-
face. If the Listen lamp is lit, the 1309 is receiving in-
structions or data from the system controller. If the Talk
lamp is lit, the 1309 is outputting data. If the SRQ lamp
is lit, the 1309 has generated a Service Request, see
section 3.7. Full details of the IEEE Interface are given in
Chapter 4.