Manual
Chapter 4
________________________________________________________________________
_______________________________________________________________________
BE1085-14
1303 Multipoint Sampler and Doser LumaSense Technologies A/S
Page 57 of 60
Table 4.7 Standardised interface jobs which perform various secondary
functions of the 1303.
Job Header Data Effect on 1303
Define_T
erminator
Decimal value of the
ASCII control character
Selects the terminator
character for interface
jobs.
Identify? No data Outputs INNOVA 1303
Output_Header EXclusive
Inclusive
Disables/enables the
1303 to output the mini-
mum code of the job
header with the appro-
priate data, in response
to the output jobs listed,
section 3.2.2.
*IDN? No data. Outputs INNO-
VA,1303,VPXXXX
*RST No data. Identical to the job Re-
set_System.
*SRE As for Ser-
vice_Request_Enable.
Identical to the job Ser-
vice_Request_Enable.
*SRE? No data Identical to the job Ser-
vice_Request_Enable?
*STB? No data Reads-
out the status byte
from the 1303.
*TST? No data Indicates if the Warning
or Error flags are set. The
1303’s condition is given
as follows:
0: Normal
1: Warning flag(s) set
-1: Error flag(s) set
The flags can then be
read-out using the Warn-
ing? or Error? jobs.