4.2.1 Set-up Jobs
These jobs install set-up parameters in the 1303. The set-up parame-
ters must be installed before starting to use the 1303. Further informa-
tion about the set-up parameters is given in section 3.3.
Table 4.1 The interface jobs which set-up the 1303.
Job Header Data Effect on 1303
Dosing_Time_Out Time interval form
10s to 3600s.
Sends the new value for
the Dosing Time-out pa-
rameter to the 1303.
Dosing_Time_Out
?
No data. Reads-out the current
Dosing Time-out value
from the 1303.
Gas_Constant Ratio of R (8314 J
mol
-1
K
-1
) and the
molecular weight M
of the tracer-gas be-
ing used.
Sends the new value for
the Characteristic Gas
Constant parameter to
the 1303.
Gas_Constant? No data. Reads-out the current
Characteristic Gas Con-
stant parameter from
the 1303.
Mol_Weight Molecular weight of
the tracer-gas being
used.
Sends the new value for
the molecular weight of
the tracer-gas being
used to the 1303.
Mol_Weight? No data. Reads-
out the molecular
weight of the tracer-gas
being used from the
1303.
Calibration_Data Doser-
nozzle number
and area of nozzle in
m
2
* 10
-9
.
Sends the calibration d
a-
ta for the stated doser
nozzle to the 1303.
Calibration_Data? Doser-nozzle num-
ber.
Reads-out the calibra-
tion data for the stated
doser nozzle from the
1303.