Manual

Chapter 3
________________________________________________________________________
_______________________________________________________________________
BE1085-14
1303 Multipoint Sampler and Doser LumaSense Technologies A/S
Page 40 of 60
ponents represented by each bit in the Status flag, and their corre-
sponding decimal values.
Table 3.3. The 1303’s Status Flag. Each bit represents a particular
1303 component; when set, the 1303 is currently using
that component.
Status Flag
Current function when
set
Bit
no.
Dec.
value
Bit name
1
1
Dosing valve 1 Dosing valve 1 open
2
2
Dosing valve 2 Dosing valve 2 open
3
4
Dosing valve 3 Dosing valve 3 open
4
8
Dosing valve 4 Dosing valve 4 open
5
16
Dosing valve 5 Dosing valve 5 open
6
32
Dosing valve 6 Dosing valve 6 open
7
64
Main dosing valve
Main dosing valve open
8
128
Dosing pump Dosing pump on
9
256
Sampling valve 1 Sampling valve 1 open
10
512
Sampling valve 2 Sampling valve 2 open
11
1024
Sampling valve 3 Sampling valve 3 open
12
2048
Sampling valve 4 Sampling valve 4 open
13
4096
Sampling valve 5 Sampling valve 5 open
14
8192
Sampling valve 6 Sampling valve 6 open
15
16384
3-way valve 3-way valve set to ana-
lyzer
16 327
68
Sampling pump Sampling pump on
Example:
STATUS?<LF> Outputs the Status flag of the 1303
as described above. Refer to Table
3.3 for the decimal values of each
bit. Refer also to the examples giv-
en below.
If the 1303 outputs 33024, using Table 3.3 shows that 33024 = 256 +
32768. Therefore, bits 9 and 16 are set. The mechanical condition of
the 1303 is: sampling valve number 1 open, and the sampling pump
running.
If the 1303 outputs 199, using Table 3.3 shows that 199 = 1 + 2 + 4 +
64 + 128. Therefore, bits 1, 2, 3, 7 and 8 are set. The mechanical con-
dition of the 1303 is: dosing valves 1, 2 and 3 open; main dosing valve
open; dosing pump running.