Technical Description
KLEIBER 740 11
Fig. 6 Connecting plug for 5-pin connecting cable
3
4
Operating and display element Meaning
3 BNC Output BNC output 0 … 10 V for connection to
storage oscilloscope
4 5-pin plug connection Power supply connection and output
[0(4) … 20] mA
Cooling water connection
At ambient temperatures of more than 40°C the pyrometer must be operated with the
associated cooling! The two connections for the connection of a cooling circuit are located
under the instrument (see
Fig. 77).
The following should be taken into account when connecting a cooling circuit:
Flow rate at least 1 l/min
Pressure: maximum 6 bar
Coolant temperature 10 °C to 35 °C
Fig. 7 Cooling water connections (tube screw connector ¼")
Cold tap water is usually sufficient for cooling. It should not be so strongly cooled that
condensation develops on the instrument. Take into account the environmental conditions
(air temperature and humidity) with respect to the cooling.