Stand-alone Settings
Monitoring Task Set-up Number = 1
Sampling = Continuous
Monitor for pre-set period = No
Compensate for Water Vapour
Interference = Yes
Compensate for cross-interference =
No
Measure Gas A: XXXXXX = Yes
Measure Water Vapour = Yes
Note: flushing is set in Environmental
Parameters, section 4.3.3
, and the
S.I.T’s are set up in the Filters setup,
section 12.3.3
1. Press SET-UP, S1 and then S1.
This brings you into the Monitoring Task branch of the Set-Up Tree
where the following text is displayed:
Up to 10 different monitoring tasks can be set-up while operating the
Monitor in Set-up mode. The parameters describing these tasks are
automatically stored in the memory of the Monitor.
2. If the number on the display is already “1” then press q to
“accept” the value.
If the number is not “1”:
a. Press
and use the direction keys to change its value to “1”.
b. Press
to accept the value.
The text which appears on the display during the next step is really
self-explanatory and it is therefore possible for you to follow why the
MONITORING TASK SET-UP NUMBER 1
PRESS ENTER TO CHANGE VALUE