50H / 100H Laser Systems Compliance With International Standards
UM-20006520DE, Rev. A Page 68
Compliance With International Standards
In compliance with international standards for laser medical equipment,
the system is equipped with the following:
Emergency Stop Button
The laser system has an emergency stop button knob that, when pushed,
immediately disables the laser system in emergency situations.
Keyswitch
Laser energy can be emitted only when the keyswitch is turned to the
Open position. The key can only be removed in the off position, and the
laser system only operates with the key in place. When treatment is
complete, always remove and secure the key to prevent unauthorized use
of the laser system.
WARNING:
Unauthorized use of this system may expose the operator/patient to poten-
tial electrical energy and laser radiation hazards.
Laser Emission Indicators
A laser emission indicator appears on the control screen to alert you that
laser energy is being emitted. During the treatment beam delivery, the
laser system emits an audible signal correlating to the pedal used.
External Door Interlock
An external door interlock outlet and plug are provided to disable the laser
system if the treatment room doors are opened while the laser system is in
Ready mode.
Protective Housing
The laser system has a protective housing that prevents unintended human
access to laser radiation. No sections of the protective housing can be
easily opened without special tools. This housing is to be opened only by
a Lumenis-certified technician.
Safety Shutter
The laser system features a safety shutter that prevents the treatment beam
from exiting the laser system. The safety shutter opens only when the
laser system is in Ready mode and the footswitch is pressed.