User's Manual
Table Of Contents
- Table of Contents
- Chapter 1: Introduction
- Chapter 2: Theory of Operation
- Chapter 3: Safety
- Chapter 4: Clinical Guide
- Chapter 5: Preparing the System for Use
- Chapter 6: Operating Instructions
- Chapter 7 - Troubleshooting and Maintenance
- Handling Error Messages and Notifications
- Troubleshooting
- Initialization Error Message Appears
- System Does Not Turn On
- Inadequate or No Aiming Beam
- No Laser Energy Emission
- “Popping” or “Tapping” Coming Sound from the Fiber Port
- Fiber Burn Back
- Unrecognized Fiber
- A Notification or Error Message Appears on the Control Panel
- System Overheats
- Message Appears: Attach an Authorized Fiber
- Message Appears: Attach fiber
- Message Appears: Attach footswitch
- Message Appears: Check footswitch
- Message Appears: Check interlock
- Message Appears: Insert debris shield
- Message Appears: No lasers
- Message Appears: Energy high
- Message Appears: Energy low
- Message Appears: Rate high
- Message Appears: Rate low
- Routine Periodic Maintenance
- Hospital/Clinic Staff Maintenance
- Professional Maintenance
- Chapter 8: System Requirements and General Information
- Appendix A: EMC Guidance and Manufacturer's Declaration
Lumenis
®
Pulse
TM
50H / 100H Laser Systems External Door Interlock Pin Assignments
UM-20006520DE, Rev. A Page 67
External Door Interlock Pin Assignments
The external door interlock is a safety feature that disables the laser
system if the treatment room doors are opened or the interlock plug is
removed while the laser system is in Ready mode.
The interlock can be set up with a remote switch, or an external switch can
be wired to the interlock plug. Plug wiring shall only be performed by a
qualified electrical professional. Total length of cable shall not exceed
five meters (16 feet).
Pin assignments are as follows:
Figure 18: External Door Interlock Pin Assignments (solder side of plug shown)
External Interlock
External Interlock RTN
Ground Shield
Locking tab with red dot
Jumper