User's Manual

Table Of Contents
Lumenis
®
Pulse
TM
50H / 100H Laser Systems Tools (Optional)
UM-20006520DE, Rev. A Page 66
Tools (Optional)
Scissors
Optical fiber stripper
Cleaving tool
Optical fiber inspection scope
SlimLine steam sterilization tray
Electrical Requirements
Refer to the System Specifications section in this manual.
Electrical Utilities
Lumenis Pulse 50H or 100H laser systems are available in several
electrical configurations. Electrical power should be setup according to
the model ordered. The line wires in the power cable shall be connected to
the building power, and the green/yellow wire must be connected to the
building ground.
The customers engineer or electrical contractor will be responsible for
ensuring that the proper electrical requirements are available at the site.
Systems Designed For Use Outside of Europe
For the 200-230 VAC configuration, supply the electrical power from a
dedicated 200, 208, or 230 VAC, single-phase, 50-60 Hz supply mains in
accordance with local codes. The socket and wall plug connection must be
rated for 30A @ 60 Hz or 30A @ 50 Hz and voltage of 200 ±10% VAC or
208 ±10% VAC or 230 ±10% VAC.
Systems Designed For Use in European Communities Under the MDD
To comply with the European Communities Medical Device Directive
93/ 42/EEC, and harmonized standards EN 60601-1 and EN 60601-2-22,
the 230 ±10%VAC configured laser must be either permanently connected
to a 32A, 230 ±10% VAC, single-phase, 50 Hz supply mains in Hard-
Wired Configurations UM-20006520EN, Rev. A Page 69 accordance with
national wiring regulations, or connected by means of a dedicated single-
phase, 32A, 230 ±10% VAC wall socket and lockable plug combination
designed to ensure the connection is “mechanically secured against
accidental loosening”. Such a connection will ensure compliance with
EN 60309 for this device. Wiring should be rated at 32A. Leakage current
for these lasers does not exceed 500 µA.