User's Manual
Table Of Contents
- Table of Contents
- Chapter 1: Introduction
- Chapter 2: Theory of Operation
- Chapter 3: Safety
- Chapter 4: Clinical Guide
- Chapter 5: Preparing the System for Use
- Chapter 6: Operating Instructions
- Chapter 7 - Troubleshooting and Maintenance
- Handling Error Messages and Notifications
- Troubleshooting
- Initialization Error Message Appears
- System Does Not Turn On
- Inadequate or No Aiming Beam
- No Laser Energy Emission
- “Popping” or “Tapping” Coming Sound from the Fiber Port
- Fiber Burn Back
- Unrecognized Fiber
- A Notification or Error Message Appears on the Control Panel
- System Overheats
- Message Appears: Attach an Authorized Fiber
- Message Appears: Attach fiber
- Message Appears: Attach footswitch
- Message Appears: Check footswitch
- Message Appears: Check interlock
- Message Appears: Insert debris shield
- Message Appears: No lasers
- Message Appears: Energy high
- Message Appears: Energy low
- Message Appears: Rate high
- Message Appears: Rate low
- Routine Periodic Maintenance
- Hospital/Clinic Staff Maintenance
- Professional Maintenance
- Chapter 8: System Requirements and General Information
- Appendix A: EMC Guidance and Manufacturer's Declaration
Lumenis
®
Pulse
TM
50H / 100H Laser Systems Change the Debris Shield Optic
UM-20006520DE, Rev. A Page 62
Change the Debris Shield Optic
1. Use snap ring pliers to remove the snap ring that secures the debris
shield optic to the body of the shield.
2. Carefully turn over the debris shield holder, and allow the debris
shield optic and spring washer to drop onto a lens cleaning tissue.
3. Open the vial containing the new debris shield optic.
CAUTION:
To avoid contamination, do not touch the surface of the debris shield optic
with your fingers.
4. Carefully turn over the vial, and allow the optic to drop onto a lens
cleaning tissue.
5. Lay the debris shield over the new debris shield optic.
6. Hold the lens cleaning tissue over the debris shield optic opening, and
slowly turn over the debris shield until the optic is facing up.
7. Place the spring washer over the debris shield optic.
8. Hold the debris shield optic and spring washer in place with the lens
cleaning tissue, and use snap ring pliers to insert the snap ring over the
spring washer. Ensure that the snap ring is in place and secure before
releasing the snap ring pliers.
Figure 17: Change the Debris Shield Optic