50H / 100H Laser Systems Emergency Stop Button Check
UM-20006520DE, Rev. A Page 59
Emergency Stop Button Check
The Emergency Stop Button is designed to disable the laser system
when pressed. To check this interlock:
1. With the system On, press down on the emergency stop button; the
system will turn itself off.
2. Restart the laser system by turning the keyswitch.
3. Press the Ready button on the LCD or footswitch to enable lasing.
If this is not the situation, discontinue use and contact Lumenis Service.
Inspect the Debris Shield
The debris shield protects the internal optical components of the laser
system from damage by a faulty or misused optical fiber.
Remove the Debris Shield
1. Turn off the laser system.
2. Locate the debris shield panel covering on the upper right-hand side of
the laser console.
3. Remove the panel covering with your fingers.
Figure 13: Locate the debris shield
Debris shield panel cover
on side of console